dc.description.abstract | La Braquiterapia 3D HDR (high doce rate) es una técnica de irradiación localizada de alta
precisión con mínimos efectos adversos, que mediante el uso de una fuente radioactiva
milimétrica permite tratar lesiones en distintas localizaciones con una mayor cantidad de
radiación emitida por unidad de tiempo utilizando un sistema de imágenes en tres dimensiones
y técnicas de planificación avanzadas.
El uso de imágenes para la planificación de tratamientos ha evolucionado de 2D a 3D,
incluyendo tomografía computada y resonancia magnética. El Hospital Oncológico de Córdoba,
realizó tratamientos de Braquiterapia intracavitaria ginecológica HDR durante seis años
consecutivos con fuente de 192Ir y basándose en planificación bidimensional (2D) mediante el
uso de películas de rayos X ortogonales. Con el objetivo de avanzar en tecnología y reducir
costos, la institución adquirió un nuevo equipo de Braquiterapia para ser utilizado con fuente de
60Co y aplicadores de materiales tales que posibilitaron la adquisición de imágenes por
Tomografía Computada y Resonancia Magnética con el fin de introducir un método de
planificación tridimensional (3D).
Entre los diversos controles que se deben realizar a las partes involucradas, como puede ser
la dosis, tiempos de exposición, planificación, etc., en la braquiterapia, el más importante es el
que involucra todo el proceso de principio a fin. Para poder llevar a cabo este trabajo se requiere de un maniquí o “phantom” lo más similar posible a un paciente para reproducir las condiciones exactas. En nuestro país no hay desarrollo comercial de maniquí específico para braquiterapia y las empresas que ofrecen dichos dispositivos son extranjeras.
En el presente trabajo se desarrolló el proceso de diseño, fabricación y prueba de un maniquí
antropomórfico para el control de calidad de principio a fin (End to end, E2E) de un tratamiento
completo de braquiterapia ginecológica HDR 3D. Una vez corroborado el buen funcionamiento
del dispositivo, con el fin de asegurar una correcta entrega del tratamiento, se realizaron
conjuntamente mediciones de QA (quality assurance) sobre la unidad HDR, la fuente radiactiva
y su conjunto dosimétrico, el equipo de adquisición de imágenes, el Sistema de Planificación de
Tratamientos (TPS), y, por último, sobre el procedimiento en general. Además, se implementó
una nueva técnica de prueba dosimétrica mediante placas radio-crómicas EBT3 (nombre de
modelo de las placas radiosensibles) que resultaron muy satisfactorias e implica un nuevo
avance en control de calidad en lo que respecta al área pública del país. Así mismo permitió
realizar una comparación con resultados de mediciones anteriores arrojando resultados muy
favorables y de alto impacto para la nueva técnica.
La incorporación de este nuevo dispositivo y las pruebas realizadas con él permitieron
actualizar el Programa de Garantía de Calidad de la institución, y establecer una base para poder implementarlo en otros servicios de Braquiterapia ginecológica HDR guiada por imágenes 3D del país. Además, cumple la función de capacitación de residentes y el perfeccionamiento de las técnicas de braquiterapia ginecológica intracavitaria. | es |
dc.description.abstract | 3D HDR (High dose rate) brachytherapy is a technique of localized high precision irradiation
with minimal adverse effects, which, by using a millimetric radioactive source, allows treating
lesions in different locations with a greater amount of radiation emitted per unit of time using a
system of three-dimensional images and advanced planning techniques.
The use of images for treatment planning has evolved from 2D to 3D, including computed
tomography and magnetic resonance imaging. The Oncological Hospital of Córdoba carried out
intracavitary gynecological HDR brachytherapy treatments for six consecutive years with a 192Ir
source based on two-dimensional (2D) planning using orthogonal X-ray films. With the aim of
advancing technology and reducing costs, the institution acquired a new brachytherapy
equipment to be used with a 60Co source and applicators of such materials that enabled the
acquisition of images by computed tomography and magnetic resonance imaging in order to
introduce a three-dimensional (3D) planning method.
Among the various controls that must be carried out on the involved parts, such as dose,
exposure times, planning, etc., in brachytherapy, the most important is the one that involves
the entire process from start to finish. To carry out this work, a mannequin or "phantom" is
required that is as similar as possible to a patient to reproduce the exact conditions. In our
country, there is no commercial development of a specific mannequin for brachytherapy, and
the companies that offer such devices are foreign.
In this work, the process of designing, manufacturing, and testing an anthropomorphic
dummy for end-to-end (E2E) quality control of a complete 3D HDR gynecological brachytherapy
treatment was developed. Once the proper functioning of the device was corroborated, in order
to ensure a correct delivery of the treatment, QA (quality assurance) measurements were jointly
performed on the HDR unit, the radioactive source and its dosimetric assembly, the imaging
acquisition equipment, the Treatment Planning System (TPS), and finally on the procedure in
general. In addition, a new dosimetric test technique was implemented using EBT3 radiochromic
plates that were very satisfactory and represent a new advance in quality control in the public
area of the country. It also allowed a comparison with previous measurement results, yielding
very favorable and high-impact results for the new technique.
The incorporation of this new device and the tests carried out with it allowed updating the
institution's Quality Assurance Program and establishing a basis for implementing it in other
services of 3D image-guided gynecological HDR brachytherapy in the country. Additionally, it
serves as training for residents and the improvement of intracavitary gynecological
brachytherapy techniques. | es |
dc.language.iso | spa | es |
dc.rights | Attribution-NonCommercial-ShareAlike 4.0 International | * |
dc.rights.uri | http://creativecommons.org/licenses/by-nc-sa/4.0/ | * |
dc.subject | Proyecto Integrador (I.Biom.) | es |
dc.subject | Braquiterapia HDR 3D | es |
dc.subject | Dosimetría | es |
dc.subject | Control de calidad | es |
dc.subject | Maniquí antropomórfico | es |
dc.subject | Test End to End | es |
dc.subject | Tomografía computada | es |
dc.subject | Planificación de tratamiento | es |
dc.title | Diseño y fabricación de maniquí prototipo antropomórfico para controles de calidad en tratamientos de Braquiterapia Ginecológica HDR 3D | es |
dc.type | bachelorThesis | es |
dc.description.fil | Fil: Levenzon Levy, Nicolas. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina. | es |
dc.description.fil | Fil: Fortuna, Gastón. Universidad Nacional de Córdoba. Facultad de Ciencias Exactas, Físicas y Naturales; Argentina. | es |